Elliptical Bimanual I/A System single-use (disposable) 21G

Elliptical Bimanual Irrigation / Aspiration System single-use, BSD 600Web14

Based on 21G, elliptical cannula system:

●  Major axis outer diameter 0.94mm, inner diameter 0.74mm
●  Minor axis outer diameter 0.62mm, inner diameter 0.42mm

1 Irrigation handpiece with curved and elliptic cannula,
proximal TWIN-PORT with diameter of 0.55mm per port

1 Aspiration handpiece with curved and elliptic cannula,
port 0.30mm, smooth abrasive (rough) tip

Product features:
●  Single use / no risk of cross contaminationWeb11
●  Lightweight, colour-coded handpieces
●  Compatible with Phaco or I/A system via standard male/female Luer connectors
●  Smooth bullet shaped tip
●  Swiss design / Swiss made

Important user information
As an elliptical system, the BSD 600 system has a higher tightness at the cannula edge. This leads to substantial less efflux. In consequence, especially the irrigation has the potential of a substantial increase of the inner eye pressure due to the specific TWIN PORT design.

To prevent any damage, e.g., any form of collapse of the eye, a reduced height of the irrigation liquid bottle is strongly recommended. Keep the pressure conditions in permanent observation and, if necessary, adjust (decrease).

Visit “Downloads” for more details and product flyers.